e-STUDIO5520C/6520C/6530C © 2008-2011 TOSHIBA TEC CORPORATION All rights reserved
REPLACEMENT OF PC BOARDS/HDD
7 - 28
7.2.5 Procedures and settings when replacing the SLG board
Note:
Before replacing the SLG board, perform the following procedure.
P.7-19 "7.2.1 Precautions when replacing PC boards"
When the SLG board has been replaced, make sure to follow the procedure below.
(1) Confirm that the power is turned OFF.
(2) Replace the SLG board.
Service Manual “7.6.18 SLG board (SLG)”
(3) Update the scanner ROM using the USB Media.
P.8-5 "8.1 Firmware Updating with USB Media"
(4) Start up with the Adjustment Mode (05).
(5) Perform “Data transfer of characteristic value of scanner / SYS board -> SLG board (05-363)”.
(6) Perform “Shading correction plate Automatic dust detection adjustment (05-349)”.
(7) Turn the power OFF.
(8) Start up with the Setting Mode (08).
(9) Check the version of the scanner ROM (08-905).
(10) Turn the power OFF.