7
ENGLISH
5.2 Potential Mechanical Damages
Damaged equipment
WARNING
Keep both hands on the upper part of the TRIOSMOVE+ when moving
it across the floor.
• Do not attempt to lift the TRIOSMOVE+ alone. When moving the system
over a doorstep or other unsmooth surface, get assistance from another
person to ensure stability.
• When moving the TRIOSMOVE+ across the floor, maintain all its wheels
in contact with the floor and ensure that they are not locked.
• Always place the scanner in its intended holder on the TRIOSMOVE+
when not in use and when moving the TRIOSMOVE+.
• Keep your feet clear of the base and wheels of the TRIOSMOVE+ when
moving it across the floor.
• Never drag the TRIOSMOVE+ across any of its cables.
• Do not place the TRIOSMOVE+ on a slope greater than 10 degrees.
5.3 Electrical Safety
Ensure a grounded/earthed connection
WARNING
To avoid the risk of electric shock, the TRIOSMOVE+ must only be connected
to a supply mains with grounding/protective earth. The TRIOSMOVE+
is equipped with a three-pin (US: prongs) grounding/earth type plug. If you
cannot insert the plug into your outlet, contact your electrician to replace the
plug or outlet, to ensure a grounded power connection. Do not try to defeat
or ignore the purpose of the grounding-type plug.
Do not access the inside of any part of the TRIOSMOVE+
WARNING
Do not access the inside of any part of the TRIOSMOVE+. Only authorized
and qualified service personnel may access the inside of any part of the
TRIOSMOVE+.
Spilled liquids
WARNING
The electrical safety of the TRIOSMOVE+ may be compromised if it
is exposed to any liquids. Do not bring liquids, such as beverages, near
the TRIOSMOVE+. Do not spill liquids, such as cleaning agent, on the
TRIOSMOVE+.
Stress on cables
CAUTION
External cables must never be subjected to pulling stress.
Condensation
CAUTION
Variations in temperature or humidity may cause water to condense inside
the system, which can cause damage. Always let the system reach room
temperature before you plug it into the power source.
If the system has been subject to major temperature or humidity changes,
wait until the system has reached room temperature before plugging it into
the power source.
If there is visible evidence of condensation, wait at least 8 hours
before plugging it into the power source, to ensure that no internal
condensation occurs.
Please see operating conditions for further information.
Disconnect the TRIOSMOVE+ before cleaning
WARNING
Disconnect the TRIOSMOVE+ from power before cleaning.
Connected USB devices
WARNING
Only bus-powered USB devices without external power supply that support
USB Full speed mode must be connected to the USB port.
Disconnecting from mains
IMPORTANT NOTICE
There is no power ON/OFF switch on the TRIOSMOVE+; therefore, the only
reliable means to disconnect the TRIOSMOVE+ from mains is to unplug the
TRIOSMOVE+ power cord. Do not position the TRIOSMOVE+ so that
it is difficult to unplug the power cord.
Please allow enough time for the system to shut down properly before
disconnecting the power cord.
5.4 Safety when Scanning
CAUTION
The TRIOS scanner is a highly precise optical instrument and must be
handled carefully. Do not drop, shake or bump the TRIOS scanner, as these
actions could cause it to break or malfunction. Follow these guidelines:
• Always return the TRIOS scanner to its holder when not in use.
• Grip the TRIOS scanner firmly during scanning and when removing/
returning it to its holder.
IMPORTANT NOTICE
The TRIOSMOVE+ will get warm during use. This is normal but not part
of any patient treatment.
5.5 Protection from Overheating
WARNING
Ventilation openings must never be blocked or obstructed. There are
openings at the back of the TRIOSMOVE+.
5.6 EMC Safety
WARNING
Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
WARNING
Use of the TRIOS system adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such
use is necessary, the TRIOS system and the other equipment should be
observed to ensure normal operation in the configuration used.
Portable RF communications equipment
WARNING
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the TRIOS system, including cables specified
by the manufacturer. Otherwise degradation of the performance of system
may occur.
5.7 Hygiene
For instructions on how to clean and disinfect or sterilize TRIOS5 scanner and other scanner
accessories, please read the safety and setup guide of the respective scanner system.
For instructions on cleaning and disinfection of the TRIOSMOVE+ go to section 6.
Always wear gloves
WARNING
To maintain a clean work environment and the safety
of the patient, wear clean examination gloves when:
• Handling the TRIOS scanner tip – changing the tip
or handling the window.
• Handling the TRIOS protection tip.
• Applying the TRIOS scanner for scanning patients.
• Touching the TRIOS system.
Prepare the TRIOSMOVE+
WARNING
In order to avoid cross contamination between patients, before each use
you must clean and disinfect relevant parts of the TRIOSMOVE+.
See chapters 6.1 and 6.2 for a description of the parts of TRIOSMOVE+ that must be cleaned
and disinfected before each use.