6253 DC Voltage Current Source/Monitor Operation Manual
4.2.8 Source and Measurement Timing
4-45
The following table shows the oscillation detection sensitivity.
4.2.8 Source and Measurement Timing
4.2.8.1 Basic Source and Measurement Timing
The following figure shows the basic timing of source and measurement.
Figure 4-13 Basic Source and Measurement Timing
• To ensure stable measurement results, set the measurement delay time Td larger than the settling
time Ts.
Tds + Ts < Td
Table 4-10 Oscillation Detection Sensitivity
Frequency 100 Hz 1 kHz 10 kHz 100 kHz 1 MHz 10 MHz
Detection sensitivity
(mVp-p)
-7050 50400 -
Tp: Repeat period for the DC and
pulse source mode, or step
period for the sweep source
mode
Tds: Source delay time
Ts: Settling time
Td: Measurement delay time
Tw: Pulse width
Tm: Measurement time
Tm = Tit + Tk
Tit: Integration time
Tk: Measurement data processing
time