6253 DC Voltage Current Source/Monitor Operation Manual
3.2 Function Description
3-3
3.2 Function Description
This section describes the panel keys and the parameter functions.
3.2.1 MENU Key (Parameter Setting)
Pressing the MENU key will display the parameter group setting screen. The MENU key goes ON.
The MENU parameters are described below:
1) Source Sets source-related parameters.
Mode Switches the source mode.
Enabled only in the Standby or Suspend status.
DC: Sets the DC source mode which generates DC voltage
or DC current.
Pulse: Sets the pulse source mode which generates pulse
voltage or pulse current.
DC Sweep: Sets the DC sweep source mode which generates
waveforms for DC voltage or current sweep.
Pulse Sweep: Sets the pulse sweep source mode which
generates waveforms for pulse voltage or current
sweep.
Pulse Base Sets the base value at pulse source.
Suspend V Sets the output voltage in the Suspend status.
Suspend Z Sets the output impedance in the Suspend status.
HiZ: The output current limit is set to 100 nA, so the output
impedance increases.
LoZ: The output current limit is set to the current limit value
at voltage source or to 3000 digits in the current range
(10000 digits in the 3 A range) at current source, so the
output impedance decreases.
Limit Input Selects the current limit HL and LL values setting.
Balance: The HL and LL values change simultaneously in
both polarities.
Individual: The HL and LL values are separately set.
Response Selects the source response.
Fast: Fast response
Slow: Slow response, but high stability to loads