6253 DC Voltage Current Source/Monitor Operation Manual
2.5.2 Pulse Source and Measurement
2-15
Operate
Press the OPR key. The key goes ON and DC voltage is output.
As VS is + 1 V and the load is 10 current of approximately 100 mA is measured.
2.5.2 Pulse Source and Measurement
An example operation of the pulse source mode is described below.
Example 2: IS: +100 mA, VL: 10 V, Load: 10
Note that the 6253 is supposed to be set as default.
5
Tp (Period) : 50 ms
Tds (Source delay time) : 0.005 ms
Td (Measurement delay time) : 4 ms
Tw (Pulse width) : 25 ms
IT (Integration time) : 1 PLC
Vp (Pulse value) : +100 mA
Vb (Base value) : 0 mA
(ex. 2)