6253 DC Voltage Current Source/Monitor Operation Manual
4.2.8 Source and Measurement Timing
4-46
4.2.8.2 Restrictions on Time Parameters
There are restrictions on setting the time parameters in relation to the others. If the time parameters are set
exceeding any of these restrictions, an error message will be displayed when the output is set to Operate
or when sweep starts, and measurement will not start.
1. Restrictions
• [Source delay time (Tds) + 94 s] < Period time (Tp) (Tds + 94 s < Tp)
• [Measurement delay time (Td) + 94 s] < Period time (Tp) (Td + 94 s < Tp)
Tp
Tds 94 s or more
Condition) Source mode: Pulse or sweep (Memory Store: other than Burst)
Tp
Td
94 s or more
Measurement
Condition) Source mode: DC, pulse or sweep (Memory Store: other than Burst)