6253 DC Voltage Current Source/Monitor Operation Manual
7.5 Calibration Procedure
7-10
Figure 7-6 Calibration Procedure (5)
[Calibration mode
setting]
[Range setting]
Enters the DMM data input mode.
DMM data
acquisition
DMM data
acquisition
DMM data
acquisition
DMM data
acquisition
Current-converted
data input
*Out-of-range setting results in a parameter error.
XD converted data
IS mode
IL-Hi mode
IL-Lo mode
Standard resistor
connection for 2 A
calibration
Within tolerance
range?
Current conversion
Current-converted value (A) =
DMM measured voltage (V) / Standard resistance value ()
Current-converted value (A) =
DMM measured voltage (V) / Standard resistance value ()
Current conversion
DMM data input
*Out-of-range setting results in a parameter error.
Checks the calibration result.
Current converted value
Current converted value
Calibration result data
save and display
Calibration result data
save and display
Current source/
limit calibration
(2 A range)
XD converted data