6253 DC Voltage Current Source/Monitor Operation Manual
3.2.1 MENU Key (Parameter Setting)
3-4
Slew Rate SW Selects the slew rate setting.
ON: Variable slew rate (SR) function ON
OFF: Variable slew rate (SR) function OFF
When the slew rate function is ON, “SR” is displayed on the
screen.
Slew Rise Sets the slew rate rise time.
Slew Fall Sets the slew rate fall time.
Slew Rate Indiv Switches the slew rate rise and fall time individual setting.
ON: The slew rate rise time and fall time are set to different
values.
OFF: The slew rate rise time and fall time are set to the same
value.
2) Sweep Mode Sets the sweep source-related parameters.
Sweep Type Selects the sweep type.
Linear: Linear sweep
Fixed: Fixed sweep
Random: Random sweep
MLinear: Multi-slope linear sweep
Log: Log sweep
Sweep Range Selects the range function for sweep source.
Auto: Sweeps in the optimum range at every step from the
start value to the stop value. *Switches to Fix in the
Burst mode.
Fix: Sweeps in the lowest range which can output any
source values from the start value to the stop value.
Sweep Reverse Switches the reverse mode (round sweep).
Single: One-way sweep from the start value and to the stop
value
Reverse: Round sweep from the start value to the stop value and
back to the start value
Repeat Count Sets the number of times sweep repeats.
When it is set to 0, sweep repeats infinitely.
When it is set between 1 and 1000 and the Sweep Reverse is set
to Reveres, each round sweep is counted as 1. *Invalid in the
Burst mode.