6253 DC Voltage Current Source/Monitor Operation Manual
8.4 Setting Time
8-12
8.4 Setting Time
Minimum pulse width: 25 s
Minimum step (repeat) time: Under fixed source/measurement range, integration time of 5 s,
the minimum measurement or source delay time, calculation function OFF,
and voltage/current measurement
Source delay time:
Period (pulse cycle):
Pulse width:
Measurement Source mode Memory mode Minimum step time
OFF
Common OFF, Normal 125 s
Common Burst 50 s
ON
DC, pulse OFF, Normal 250 s
*6)
Sweep
Burst 50 s
OFF, Normal 250 s
*6)
*6) 500 s when the calculation function is ON
Setting range Resolution
*7)
Setting accuracy
0.005 ms to 60.000 ms 1 s
(0.1 % + 10 s)
60.01 ms to 600.00 ms 10 s
600.1 ms to 6000.0 ms 100 s
6001 ms to 59997 ms 1 ms
Setting range Resolution
*7)
Setting accuracy
0.050 ms to 60.000 ms 1 s
(0.1 % + 10 s)
60.01 ms to 600.00 ms 10 s
600.1 ms to 6000.0 ms 100 s
6001 ms to 60000 ms 1 ms
*7) The setting resolution is determined by the period time resolution.
Setting range Resolution
*7)
Setting accuracy
0.025 ms to 60.000 ms 1 s
(0.1 % + 10 s)
60.01 ms to 600.00 ms 10 s
600.1 ms to 6000.0 ms 100 s
6001 ms to 59997 ms 1 ms