2D-LC User Guide 21
2 Concepts of 2D-LC
Multiple Heart-Cutting and High Resolution Sampling 2D-LC
Peaks that are cut out and stored during a run are analyzed consecutively in the
second dimension, even when the first dimension is still running. To avoid
carry-over the peaks are analyzed in reverse order of storage in a single Multiple
Heart-Cutting Valve.
D
1
D
2