2D-LC User Guide 293
10 Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
To upgrade/downgrade the module’s firmware carry out the following steps:
1 Download the required module firmware, the latest Lab Advisor software and
the documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module
a Select the folder on the hard drive where the Firmware package is stored.
b Connect the Lab Advisor Software to your 2D-LC instrument.
Figure 153 Firmware Update
c Press the Lock button.
The system is locked.
d Select the required firmware version for the Resident and Main Firmware.
e To update the firmware of the instrument, press the Update button.
This will require some time.
Do not interrupt the power supply of the device and the PC during this procedure.