Agilent 4150 TapeStation System Manual 107
Good Measurement Practices
7
Quantitative range
Quantitative range
To achieve accurate results the samples used must be within the quantitative
range as stated in the respective assay Quick Guide. Signal of samples with
concentrations below then the lower limit of the quantitative range or below
the limit of detection might not be visible. Signal of samples with
concentrations higher than the upper limit of the quantitative range might be
significantly higher than the marker peaks. Markers might not get assigned in
this case (Table 3 on page 64).
Figure 34 Example of missing marker peak due to too concentrated sample in the High
Sensitivity D1000 assay in the Gel view (Left) and Electropherogram view
(Right) of the TapeStation Analysis software.