32 Agilent 4150 TapeStation System Manual
3
Product Description
Physical Specification
Physical Specification
Table 2 Physical Specifications
Type Specification
Weight 17.8 kg (39.2 lbs)
Dimensions (w×d×h) 254 x 510 x 415 mm (9.6 x 20.1 x 16.3 inches)
Line voltage 100 – 240 VAC ± 10 %
Line frequency 50 or 60 Hz ± 5 %
Power consumption 50 VA
Interface USB2 (Instrument: USB-B; PC: USB-A)
Ambient operating temperature 10 – 40 °C (50 – 104 °F)
Operating humidity 15 – 80 %, non-condensing
Storage temperature -40 to 70 °C (-40 to 158 °F)
Operating altitude Up to 3000 m (9842 ft)
Safety standards IEC, EN, CSA, UL, Installation category II,
Pollution degree 2
For indoor use only.
ISM Classification ISM Group 1 Class A
According to CISPR 11
Sound pressure <70 dB (A)
Assay specific temperature range 12 – 37 °C (54 – 99 °F) for D1000 ScreenTape
17 – 37 °C (63 – 99 °F) for High Sensitivity D1000 ScreenTape
15 – 30 °C (59 – 86 °F) for Genomic DNA ScreenTape
14 – 30 °C (57 – 86 °F) for RNA and High Sensitivity RNA ScreenTape
15 – 30 °C (59 – 86 °F) for D5000 ScreenTape
15 – 30 °C (59 – 86 °F) for High Sensitivity D5000 ScreenTape
16 – 30 °C (61 – 86 °F) for Cell-free DNA ScreenTape
This is an ISM Class A product intended for use in industrial environment. In a domestic
environment, this product may cause radio interference, in which case the user may be
required to take adequate measures.
The assay specific temperature range is measured at the ScreenTape nest. Choose the
location of the instrument and the ambient conditions to fit to the working range of the
assay else the start of runs will be disallowed.