Agilent 4150 TapeStation System Manual 133
Maintenance
9
Electrode Cartridge
Electrode Cartridge Exchange Process Detail
Electrode Cartridge exchange process detail
The Electrode Cartridge change process includes several background tasks
such as:
electronic deregistration of the old cartridge,
electronic registration of the new cartridge
validation of information stored on the internal memory
short functionality test with the TapeStation Test Tape
After an Electrode Cartridge Change a banner shows results of the outcome.
Users are subsequently guided to execute a System Validation test in any case.
This additional system validation is not mandatory to continue working with
the instrument but is recommended for good laboratory practice reasons.
The Electrode Cartridge Change is successful if the following criteria are
fulfilled:
The inserted Electrode Cartridge was successfully registered
The date of first use is the current day
No samples were analyzed with it
The electrical functionality test with the TapeStation Test Tape passed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4150 TapeStation system and is the answer not in the manual?

Agilent Technologies 4150 TapeStation system Specifications

General IconGeneral
BrandAgilent Technologies
Model4150 TapeStation system
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals