110 Agilent 4150 TapeStation System Manual
7
Good Measurement Practices
Maximum Buffer Concentration in Sample
Maximum Buffer Concentration in Sample
The Maximum buffer concentration in sample describes the maximum
concentration of additional salt, namely ionic components in sample.
ScreenTape assays are designed to compensate for negative effects salt
introduces during electrophoresis. The buffer in which the nucleic acid sample
is provided (that is, the sample matrix) must not exceed these maximum salt
concentrations. Otherwise the performance is significantly affected. The
Maximum buffer concentration in sample can be found in the respective
Quick Guide (Table 3 on page 64).