188 Agilent 4150 TapeStation System Manual
10
Troubleshooting
Troubleshooting DNA Applications
Unexpected Migration Profile
There are a number of issues which can affect migration and sample peak
profile. If the issues below do not describe your analysis file, please contact
your support representative.
Table 10 DNA - probable causes for unexpected migration profile
Problem Likely cause Explanation Solution
The gel image looks
distorted
Incorrect Marker
peaks picked up
Depending on the
sample pattern wrong
peaks can get
assigned as marker if
running very close to
the actual marker.
Use the Aligned button in the TapeStation
Analysis software to detect the correct
markers. Compare to aligned/unaligned data
of good and questionable samples; and then
assign correct markers by right-clicking.
See the Agilent Information Center for more
information.
Ensure that all sample peaks are within the
recommended sizing range for the application,
see “Peak Integration” on page 101.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4150 TapeStation system and is the answer not in the manual?

Agilent Technologies 4150 TapeStation system Specifications

General IconGeneral
BrandAgilent Technologies
Model4150 TapeStation system
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals