Agilent 4150 TapeStation System Manual 127
Maintenance
9
Exchange the Piercing Needle
Exchange the Piercing Needle
Beyond the descriptions in this user guide you will also find a detailed
description within the Agilent Information Center (see “Agilent Information
Center” on page 60).
1 Prepare instrument by selecting Tools > Maintenance > Change needle from
the TapeStation Controller software. The exchange process opens the
ScreenTape nest to allow insertion of the Needle Change Cartridge.
2 Close Lid. Click Proceed.
Parts required p/n Description
5067-5783 Needle Change Cartridge
CAUTION
Analyses fail
Turning the instrument off during the calibration procedure after the needle exchange
will result in a fail of all subsequent analysis. Turning the instrument off during
calibration will require a different new needle change cartridge.
Do not turn the instrument off during the calibration procedure after needle
exchange.
CAUTION
Damage to the cartridge or nest
The Needle Change Cartridge cannot be inserted into the ScreenTape nest in normal
operation mode. It is required to start the needle exchange process by software first,
this will open the ScreenTape nest.
Do not use excessive force when inserting the Needle Change Cartridge into the
ScreenTape nest.
Insert Needle Change Cartridge to the ScreenTape nest only when requested by the
software dialogue.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4150 TapeStation system and is the answer not in the manual?

Agilent Technologies 4150 TapeStation system Specifications

General IconGeneral
BrandAgilent Technologies
Model4150 TapeStation system
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals