120 Agilent 4150 TapeStation System Manual
8
Instrument Diagnosis
Tests
Executing System Verification Tests
1 Turn on the laptop. Let the operating system start entirely.
2 Power on the TapeStation instrument with the instrument lid closed. Let it
initialize entirely.
3 Start the TapeStation Controller software and allow the software to detect
the instrument.
4 Select System Diagnostics > Run Tests.
5 Select one of the three test options, see “System Diagnostics on page 57,
for example the System Verification tests.
6 Select the default tests or omit tests according to your needs. Individual
tests can be checked or unchecked.
7 Click Start and follow the on-screen instructions closely.
NOTE
Some individual tests are depended on others. When a test is deselected, any dependent
tests are also deselected.