26 Agilent 4150 TapeStation System Manual
3
Product Description
Product Descriptions
2 Magnet
Figure 5 Position of magnet in the instrument lid
WARNING
Heart pacemakers
Magnets could affect the functioning of pacemakers and implanted heart
defibrillators.
A pacemaker could switch into test mode and cause illness.
A heart defibrillator may stop working.
Bearers of heart pacemakers or implanted defibrillators must stay off at least
55 mm from the magnets.
CAUTION
Magnetic fields
Magnets produce a far-reaching, strong magnetic field.
You can damage for example televisions, laptops, computer harddisks, credit cards,
magnetic cards may be damaged as well.
Keep magnets at least 25 mm away from devices and objects that could be
damaged by strong magnetic fields.