Installation
4
Site Requirements
Missing electrical ground
Electrical shock
➔ If your product is provided with a grounding type power plug, the instrument
chassis and cover must be connected to an electrical ground to minimize shock
hazard.
➔ The ground pin must be firmly connected to an electrical ground (safety ground)
terminal at the power outlet. Any interruption of the protective (grounding)
conductor or disconnection of the protective earth terminal will cause a potential
shock hazard that could result in personal injury.
Instrument covers removed
Electrical shock
➔ Do Not Remove the Instrument Cover
➔ Only Agilent authorized personnel are allowed to remove instrument covers.
Always disconnect the power cables and any external circuits before removing the
instrument cover.
Damage to the instrument
Personal injury (for example electrical shock, intoxication)
➔ Instruments that appear damaged or defective should be made inoperative and
secured against unintended operation until they can be repaired by qualified
service personnel.