Agilent 4150 TapeStation System Manual 37
Installation
4
Bench Space Required
Allow approximately 75 mm (3 inches) of clear space around the instrument
for air circulation and electrical connections.
Performance can be affected by sources of heat and cold, for example direct
sunlight, heating/cooling from air conditioning outlets, drafts and/or
vibrations.
A computer system is required to control your instrument. Account for
respective electrical outlets. Two plug sockets must be available near the
allocated bench space for the instrument itself and a laptop.
PP
PP
PP
PP
NOTE
Additional space of approximately 25 cm is required on top of the listed height in order to
open the TapeStation lid. Please avoid locations with low overhanging shelves.