Standalone Control
Setting the run parameters
93
• Number of sample washes—There are two switches for 0, 2, 6, or 10 
sample washes
• On-column injection mode—There is one no/yes switch.
• Number of solvent washes—There two switches for 0, 2, 6, or 10 solvent 
washes.
• Position of first vial—If the tray is installed, the first vial must be in tray 
position 1. Without a tray, the first vial must be in turret position 1.
• Position of last vial—If the tray is installed, it starts with vial position 1 
and continues until it does not find a vial in the next position. Without a 
tray and with more than one vial, the last vial is identified by the setting 
made on the rotary switch located at the top of the tower.
The following run parameters are either preset or not available:
• Number of pumps—This is preset to six pumps.
• Viscosity—There is 0 second delay at the top of the pump and injection 
strokes.
• Look for priority samples—This is not available with standalone control. 
• Position—The injector position is based on the position of the injector 
cable connection in the controller module.