Operation
6890 Series GC Control
80
Cancel the store, and return to the STORED SEQUENCES status table.
Sequences can also be stored from within the STORED SEQUENCES status
setpoint table by scrolling the cursor to the appropriate sequence number and
pressing the [Store] key.
Starting/running a sequence
Press the [Seq Control] key, scroll to the Start sequence line, and press
[Enter].
Pressing [Enter] changes the sequence status to Running.
If a sequence is started and the instrument is not ready (due to oven
temperature, equilibration times, etc.), the sequence will not start until all
instrument setpoints are ready.