EasyManua.ls Logo

Agilent Technologies 7673 - Filling and Placing the Bottles

Agilent Technologies 7673
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing for Operation
Preparing the solvent and waste bottles
37
For many common solvents, the rate of diffusion out of the bottle is less
with a diffusion cap than with a septum that has been punctured with a
standard syringe needle.
Figure 27. 4-ml bottle used for solvent and waste
Filling and placing the bottles
Before each sequence or group of sequences, prepare your solvent and waste
bottles by doing the following:
1. Rinse and fill each solvent bottle with 4 to 4.5 ml of fresh solvent. The
liquid level should be near the shoulder of the bottle. If the solvent bottle is
filled with 4.5 ml of solvent, the syringe can reach about 2 ml or about 250
washes for a 10-µl syringe.
Figure 28. Shows position of the syringe tip when withdrawing solvent
Diffusion cap
Agilent Technologies recommends
using a diffusion cap instead of a
septum to reduce solvent
contamination and evaporation.
4.5 ml maximum
2.5 ml minimum
18.5 mm

Table of Contents

Related product manuals