Preparing for Operation
Preparing the solvent and waste bottles
38
2. Empty and rinse each waste bottle. The syringe can dispense about 4 ml of
waste into the waste bottle or about 500 washes for a 10-µl syringe.
Figure 29. Shows position of syringe tip when dispensing waste
3. Place the bottles in the appropriate positions on the injector turret
according to the table on page 40. The positions are labeled on the turret,
Solvent A, Waste A, Solvent B, and Waste B.
Caution Do not refill a solvent bottle that still has solvent left in the bottle. The solvent
from the last analysis may be contaminated.
Figure 30 shows an example of the turret positions used with and without a
tray.
If a tray is installed, place four bottles in turret positions solvent A, waste
A, solvent B, and waste B.
4 ml maximum