Preventive Maintenance and Troubleshooting
Fault light symptoms
135
Syringe error
Probable causes
The syringe carriage motor has a problem.
The syringe carriage cannot find its home position.
The syringe carriage sensor is inoperable.
The improper injection mode was used.
For information on the syringe carriage home positions, see the Glossary.
Suggested actions
1. Ensure the syringe is installed correctly. For more information, see
“Selecting and installing syringes” on page 49.
2. Ensure the syringe meets specifications.
3. Check the on-column injection mode parameter. If the parameter was set
incorrectly for on-column injection, check for damage to the inlet liner and
column.
4. If the syringe needle is bent, also check the list of actions on page 119.
5. Press the STOP button on the injector to clear the fault indicators. Restart
the sequence.
6. If the error occurs again, obtain Agilent service.