EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7673

Agilent Technologies 7673 User Manual

Agilent Technologies 7673
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Preparing for Operation
Preparing the solvent and waste bottles
40
Caution When a tray is installed, place a waste bottle in both waste position A and
waste position B. The injector alternates dispensing waste between the two
positions. With the 6890 Series GC, you can choose A, B, or both.
Do you need to read further?
The volumes of the solvent and waste bottles determine the number of sample
vials that you can run at one time. If your application requires more than the
maximum number of washes listed in Figure 31, you must read the next section
entitled, “Estimating the maximum number of sample vials.” The total number
of washes includes all solvent (pre- and post-injection) and sample washes
using the default needle depth. This is critical to your solvent supply if solvent
prewashes are not part of your solvent needed estimation.
System Solvent Waste
Standalone With a tray
Without a tray
A & B
A
A & B
A
3365/5890 With a tray
Without a tray
A, B, or both
A, B, or both
A & B
A
3366/5890 With a tray
Without a tray
A, B, or both
A
A & B
A
3396/6890 With a tray
Without a tray
A, B, or both
A, B, or both
A, B or both
A
Multitechnique
ChemStation/5890
With a tray
Without a tray
A, B, or both
A, B, or both
A & B
A
Multitechnique
ChemStation/6890
With a tray
Without a tray
A, B, or both
A, B, or both
A, B or both
A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7673 and is the answer not in the manual?

Agilent Technologies 7673 Specifications

General IconGeneral
BrandAgilent Technologies
Model7673
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals