Preventive Maintenance and Troubleshooting
Fault light symptoms
139
Bottle in gripper
Probable cause
• The sample vial was not delivered properly and stayed in the tray gripper.
Suggested actions
1. Remove the vial and return it to its position in the tray.
2. Ensure that the tray quadrants are snapped into place.
3. Ensure that the injector is plugged into the correct connector on the back 
of the controller (front or rear).
4. Check the “deliver to” location for the vial, and verify that the location is 
empty and free from obstructions.
5. Press the STOP button on the injector to clear the fault indicators. Restart 
the sequence.
6. If the error occurs again, obtain Agilent service.