Preventive Maintenance and Troubleshooting
Fault light symptoms
138
Incomplete injection
Probable causes
There was an incomplete injection.
The plunger or syringe carrier is operating incorrectly during injection.
Suggested actions
1. If the syringe needle is bent, check the list of actions on page 119.
2. Remove the syringe from the injector, and check the plunger for stickiness
or binding. Replace the syringe if necessary.
3. Press the STOP button on the injector to clear the fault indicators. Restart
the sequence.
4. If the error occurs again, obtain Agilent service.