Standalone Control
Setting the run parameters
97
Caution The slow setting is only recommended for a few specific splitless and cool on-
column applications. For most applications, set this switch to “fast.” A setting
of “slow” for other applications may cause peak area discrimination and
reduced quantitative precision.
Number of injections per vial
Set the number of injections per vial to 1, 2, 3, or 4. The following table shows
the number of injections for each of the switch settings.
Figure 55. Switches for setting the number of injections from each sample vial
1
2
3
Switch Setting
Both up
Left up, right down
Both down
Left down, right up 4
Injections/Sample Vial
Set for four injections per vial
injections
#/vial
1
2
3
4