Operation
Checklist
68
Checklist
Use this checklist to make sure the sampler is ready before you begin.
3396
Sample vials are half full
Cap centered, no wrinkles, septum is flat
Sample inserts and vials match the run parameters
Tray quadrants snapped in place
4.5 ml of fresh solvent in each solvent bottle
Waste bottles are clean and empty
Two waste bottles (with tray)
Limit to number of sample vials
New or clean syringe
Correct syringe design and size
Plunger carrier holds plunger button
Needle is aligned with septum retainer nut
Syringe is rinsed with solvent
Liner is clean and deactivated
Liner is correct type for injection technique
Septum type is correct
Septum is less than 200 injections old
Correct septum nut
Injector run parameters are set correctly
Injection mode matches type of inlet
Number of injections per vial is less than five
Injection volume is not greater than 2 µl for splitless
inlet