Operation
What happens during a run?
67
The injector waits until the GC sends a ready signal to continue. Steps 1, 2, and
3 through 9 are repeated according to the “injections per vial.”
With a tray, the steps are repeated until the tray returns the last sample to the
“position of last vial.” Without a tray, the steps are repeated according to the
rotary switch setting.
If the “priority sample” feature is turned on, the tray gripper looks for a sample
in tray quadrant position 100 after each vial. After a priority sample is run, the
tray returns it to tray quadrant position 99.