Standalone Control
Running the samples
105
For more information, see chapter 5, “Preventive Maintenance and
Troubleshooting.”
The sampler does not remember which sample vial was associated with the
interruption. When you restart an interrupted sequence, the sampler returns
any vials that were left in the injector turrets and begins a new sequence.
To restart the sequence from the point of interruption:
1. Identify the last sample vial that was run successfully. Check the vial
number of the last successful chromatogram.
2. Reposition the sample vials so that the next bottle you want to run is in
position 1.
3. Press the stop button to clear any faults.
4. Press the start button.