Standalone Control
Example of standalone control setup
110
3. Set the number of injections per sample vial to one.
4. Set the number of sample washes per sample vial to two.
5. The number of pumps is preset to six.
6. The viscosity delay is preset to zero.
7. Set the sample size (injection volume) to 1 µl.
8. Set the number of solvent A washes to two.
• With a tray, the two solvent washes are split so that the injector 
performs one solvent A and one solvent B wash.
• Without a tray, the injector performs two solvent A washes.
injections
#/ vial
sample
pre-wash
Left switch up,
right switch down
sample
size
Both switches up