EasyManua.ls Logo

Agilent Technologies 7673 - Page 65

Agilent Technologies 7673
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing for Operation
Adapting for cool on-column injection
56
Slows the plunger speed to 1/18 the normal speed
Complete the following:
1. Install an on-column syringe. Agilent Technologies recommends syringes
with a 23/26 gauge tapered-tip needle for injections onto 530-µ columns
and 32/26 gauge for 320- and 250-µ columns.
2. If the injector is controlled by an external device, such as an integrator or
a workstation, change the on-column parameter to ON or YES.
If the injector is controlled by a 6890 GC, this will be set automatically
when the GC configures for the on-column inlet.
If the injector is operated with standalone control, change the on-column
switch to the YES, or down, position. For more information on the on-
column switch, see “On-column injection mode” in chapter 4.
3. The inlet must be configured for cool on-column injection. Install the
appropriate needle guide, septum, spring, and insert for the GC.
If you have an on-column inlet on a 5890A GC with a cooling tower
(part no. 19245-20540), replace it with a septum nut base assembly
(part no. 19245-80520).
If you are performing automated injection onto 320-µm and 250-µm
columns, replace the standard septum nut base assembly (part no.
19245-80520) with the small septum nut base assembly (part no.
07673-61330)
4. For injecting onto 320-µm or 250-µm columns, see “Performing cool on-
column injection onto 250-µm and 320-µm columns” in chapter 6.

Table of Contents

Related product manuals