Operation
3396 integrator with a 5890 GC
84
10. Set the viscosity delay to zero.
VISCOSITY --> Press [0] [ENTER].
11. Set the sample size (injection volume) to 1 ml.
VOLUME --> Press [1] [ENTER].
12. Set the injection speed to fast.
SLOW INJECTION --> Press [0] [ENTER].
13. Set the number of solvent A washes to one.
# OF SOLVENT A WASHES --> Press [1] [ENTER].
14. With a tray, set the number of solvent B washes to one.
# OF SOLVENT B WASHES --> Press [1] [ENTER].
Without a tray, skip to the next instruction.
15. Turn the priority sample feature off.
PRIORITY SAMPLE (1=YES) --> Press [0] [ENTER].
16. Set the injection mode. CAPILLARY ON-COLUMN
Normal for capillary or packed inlet.
Press [0] [ENTER].
or
On-column for an on-column inlet.
Press [1] [ENTER].