EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8563E

Agilent Technologies 8563E User Manual

Agilent Technologies 8563E
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #440 background imageLoading...
Page #440 background image
440 Chapter7
Language Reference
CLRW Clear Write
CLRW Clear Write
Syntax
Figure 7-79 CLRW Syntax
Description
The CLRW command sets the chosen trace to clear-write mode. This
mode sets each element of the chosen trace to the bottom-screen value;
then new data from the detector is put in the trace with each sweep.
Example
10 OUTPUT 718;"IP;"
20 OUTPUT 718;"BLANK TRA;CLRW TRB;"
30 END

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8563E and is the answer not in the manual?

Agilent Technologies 8563E Specifications

General IconGeneral
BrandAgilent Technologies
Model8563E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals