Chapter 7 571
Language Reference
RBR Resolution Bandwidth to Span Ratio
RBR Resolution Bandwidth to Span Ratio
Syntax
Figure 7-219 RBR Syntax
Description
The RBR command specifies the coupling ratio between the resolution
bandwidth and the frequency span. When the frequency span is
changed, the resolution bandwidth is changed to satisfy the selected
ratio.
Parameters
number real from 0.002 to 0.10.
UP or DN increments in a 1, 2, 5 sequence.
Preset State
0.011
Query Response
Figure 7-220 RBR Query Response