EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8563E

Agilent Technologies 8563E User Manual

Agilent Technologies 8563E
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #481 background imageLoading...
Page #481 background image
Chapter 7 481
Language Reference
GATE Gate
GATE Gate
Syntax
Figure 7-122 GATE Syntax
Description
The GATE command turns on or off the time-gating function. When the
time-gating function is turned on, the spectrum analyzer activates the
time gate circuitry according to the parameters controlled by gate
length (GL), gate delay (GD), and the gate trigger input.
Preset State
Off
Query Response
A "0" is returned if the time-gate function is off, a "1" is returned if the
time-gate function is on.