606 Chapter7
Language Reference
SRCPSTP Source Power Step
SRCPSTP Source Power Step
Syntax
Figure 7-253 SRCPSTP Syntax
Description
The SRCPSTP command sets the step size of the source power level,
source power offset, and power-sweep range functions. This function is
available only with an 8560E/EC Option 002.
Parameters
number real from 0.1 dB to 12.75 dB; 0.05 dB resolution via
GPIB.
UP or DN 0.1 dB steps.
Preset State
1.0 dB
Query Response
Figure 7-254 SRCPSTP Query Response