508 Chapter7
Language Reference
MKBW Marker Bandwidth
MKBW Marker Bandwidth
Syntax
Figure 7-156 MKBW Syntax
Description
When used remotely, the MKBW command finds the signal bandwidth
at the power level in dB below the on-screen marker (if a marker is
present) or the signal peak (if no on-screen marker is present). When
the command is used manually, a peak search is automatically
performed, and the bandwidth of the largest signal on-screen is
displayed in the message area.
Parameter
integer any valid negative integer (0 to the amplitude of the
noise floor).
Example
10 OUTPUT 718;"IP;"
20 OUTPUT 718;"CF 300MHZ;SP 100MHZ;SNGLS;"
30 OUTPUT 718;"TS;MKPK HI;"
40 OUTPUT 718;"MKBW -3,?;"
50 ENTER 718;Mkbw
60 END