EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8563E

Agilent Technologies 8563E User Manual

Agilent Technologies 8563E
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #604 background imageLoading...
Page #604 background image
604 Chapter7
Language Reference
SRCPOFS Source Power Offset
SRCPOFS Source Power Offset
Syntax
Figure 7-251 SRCPOFS Syntax
Description
The SRCPOFS command offsets the displayed power of the built-in
tracking generator. This function can be used to take into account
system losses (for example, cable loss) or gains (for example,
preamplifier gain) reflecting the actual power delivered to the device
under test. SRCPOFS is available only with an 8560E/EC Option 002.
Parameters
number real from 100 dB to +100 dB.
UP or DN increments in steps equal to the value set by SRCPSTP.
Preset State
0.0 dB

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8563E and is the answer not in the manual?

Agilent Technologies 8563E Specifications

General IconGeneral
BrandAgilent Technologies
Model8563E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals