576 Chapter7
Language Reference
RCLT Recall Trace
RCLT Recall Trace
Syntax
Figure 7-223 RCLT Syntax
Description
The RCLT command recalls previously saved trace data to the display.
See SAVET.
Parameters
TRA recalls the trace data to trace A.
TRB recalls the trace data to trace B.
number integer from 0 to 7. If a Agilent 85620A mass memory
module is attached to the spectrum analyzer, trace
registers 5, 6, and 7 are used by the Agilent 85620A
mass memory module. Numbers less than zero default
to zero; numbers greater than seven default to seven.
Example
10 OUTPUT 718;"IP;CF 300MHZ;SP 20MHZ;"
20 OUTPUT 718;"SAVET TRA,7;"
30 OUTPUT 718;"IP;"
40 OUTPUT 718;"RCLT TRB,7;"
50 END