Chapter 9 677
If You Have a Problem
Procedures
Plotting and Printing Directly
1. The printer or plotter must be connected to the spectrum analyzer
GPIB bus.
2. No other controller can be on the bus when doing direct plotter or
printer dumps.
3. The printer or plotter cannot be set to a listen only mode, it must be
addressable.
4. Printing is selected by pressing CONFIG and COPY DEV PRNT PLT so
that PRNT is underlined. Plotting is selected by pressing the key so
that PLT is underlined.
5. The factory default GPIB address for a printer is 01. For a plotter, it
is 05. This address can be changed by pressing
CONFIG and either
PRINTER CONFIG or PLOTTER CONFIG to access the address
functions.