692 Chapter9
If You Have a Problem
Error Messages
Error codes 200 through 299 relate to ADC hardware/firmware
failures. Instrument service is required.
ERR 200 SYSTEM ADC Driver/ADC hardware/firmware interaction;
check for other errors.
ERR 201 SYSTEM ADC Controller/ADC hardware/firmware
interaction; check for other errors.
ERR 202 FADC CAL Linear offset search failed.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 203 FADC CAL Log offset search failed.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 204 FADC CAL Log 2× expand offset search failed.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 205 FADC CAL Linear slope calibration is out of range.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 206 FADC CAL Log slope calibration is out of range.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 207 FADC CAL Log 2× expand slope calibration is out of range.
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
ERR 250 OUTOF RG ADC input is outside of ADC range.
ERR 251 NO IRQ Microprocessor not receiving interrupt from ADC.