Chapter 9 695
If You Have a Problem
Error Messages
ERR 356 SPAC CAL Sweep data problem finding "bucket 1" of the
span accuracy calibration sweep.
ERR 357 SPAC CAL Cannot find the "x" intersection for "bucket 1"
of the span accuracy calibration sweep.
ERR 358 SPAC CAL Cannot find "bucket 2" of the span accuracy
calibration sweep.
ERR 359 SPAC CAL Cannot find the "x" intersection for "bucket 2"
of the span accuracy calibration sweep.
ERR 360 SPAC CAL The start bucket correction is out of range.
ERR 361 SPAC CAL The percent of span correction is out of range.
Error codes 400 through 599 relate to IF failures. Instrument
service is required These errors are generated when the automatic
IF adjustment routine detects a fault. These errors are designed to
assist service personnel in IF problem isolation, however, an
instrument will function if the suspect functional parameters are not
required. This routine adjusts amplitude parameters first, then
resolution bandwidths in this sequence: 300 kHz, 1 MHz, 2 MHz, 100
kHz, 30 kHz, 10 kHz, 3 kHz, 1 kHz, 300 Hz, 100 Hz, 30 Hz, 10 Hz, 3 Hz,
and 1 Hz. The routine restarts from the beginning if a fault is detected.
Resolution bandwidth filters adjusted after the routine begins and
before the fault is detected should be acceptable; resolution bandwidth
filters adjusted later in the sequence are suspect. If the fault is detected
before the first resolution bandwidth (300 kHz) is adjusted, all IF
performance is suspect.
ERR 400 AMPL<300 Unable to adjust less than 300 Hz RES BW
amplitude.
ERR 401 AMPL 300 Unable to adjust amplitude of 300 Hz RES BW.
ERR 402 AMPL 1K Unable to adjust amplitude of 1 kHz RES BW.
ERR 403 AMPL 3K Unable to adjust amplitude of 3 kHz RES BW.
ERR 404 AMPL 10K Unable to adjust amplitude of 10 kHz RES BW.
ERR 405 RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 406 RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 407 RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 408 RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 409 RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 410 RBW 10K Unable to adjust 10 kHz RES BW pole 2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8563E and is the answer not in the manual?

Agilent Technologies 8563E Specifications

General IconGeneral
BrandAgilent Technologies
Model8563E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals