EasyManua.ls Logo

Agilent Technologies InfinityLab LC Series - Page 127

Agilent Technologies InfinityLab LC Series
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
G4208A User’s Guide 127
4 Running Multiple-Vial Analyses
Re-calibrating With Multiple Groups of Standards
4 Move to the End of sequence list, select the Wizard and enter:
5 Select Calibration to display the Calibration Settings screen.
6 Change the settings according to Figure 84 on page 127.
Figure 84 Sequence Calibration Wizard
7 You may press Preview to view the result.
8 Select Done to accept entries.
9 Move to the End of sequence list and press Insert.
10 Select Method and then move to METH2 and press Load.
11 Move to the End of sequence list, select the Wizard and enter:
12 Select Calibration to display the Calibration Settings screen.
Vial Range 7 to 12
#Inj. 1
Vial Range 13 to 21
#Inj. 1
NOTE
You have to insert METH2/METH3 before/after each calibration sample line(s).
Use copy/paste or use insert/parameter/stoptime and only one method.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Related product manuals