G4208A User’s Guide 146
5 Maintenance and Repair
Firmware Updates
Downgrade from B.xx to A.05.1x
1 When the Agilent 1100/1200 series modules are on a firmware platform
A.06.xx, downgrade the Instant Pilot to A.05.1x first.
During the downgrade process the screen is black.
2 After boot of the Instant Pilot,
the Instant Pilot shows “Scanning System ...
the LAN/RS- 232 Update Tool 2.10 indicates “100% updated” but does not
display the result dialog.
3 Press Cancel (red cross button) on the LAN/RS- 232 Update Tool 2.10 to stop
the Update process.
The Instant Pilot is no longer listed in the tree view of the Update Tool.
4 Downgrade all 1100/1200 modules together to the target firmware
(A.05.11/13 or A.05.09/10).
When finished,
the LAN/RS- 232 Update Tool 2.10 shows all 1100/1200 modules without
the Instant Pilot.
the Instant Pilot shows all modules of the system in the Welcome screen.
5 Close the connection and the LAN/RS- 232 Update Tool 2.10.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals