Configuring the Heartstream XL
10-12 Setting Up and Configuring the Heartstream XL
7DEOH0DQXDO6HWWLQJV
7DEOH(&*)LOWHU6HWWLQJV
Parameter Description Choices
6\QF$IWHU6KRFN Determines if the Sync function stays on
after a synchronized shock is delivered.
On, Off
'LVSOD\6KRFN
&RXQWHU
Determines if the number of shocks
delivered appears on the display during
an event.
On, Off
'LVSOD\,QFLGHQW
7LPHU
Determines if the elapsed time appears on
the display during an event.
On, Off
Item Description Setting Choices
$&/LQH)LOWHU Selects the setting used to filter out
AC line noise.
60 Hz, 50 Hz
3DGV(&*IRU
'LVSOD\
Selects the display filter frequency
for the therapy cable attached.
Monitor (.15-40Hz),
EMS (1-30 Hz)
3DGV(&*IRU
3ULQWHU
Selects the printer filter frequency
for the therapy cable attached.
Monitor (.15-40Hz),
EMS (1-30 Hz)
/HDGV(&*IRU
'LVSOD\
Selects the display filter frequency
for the monitoring electrodes ECG.
Monitor (.15-40Hz),
EMS (1-30 Hz)
/HDGV(&*IRU
3ULQWHU
Selects the printer filter frequency
for the monitoring electrodes ECG.
Diag (.05 - 150 Hz),
EMS (1 - 30 Hz,
Monitor (.15 - 40 Hz)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies M4735A and is the answer not in the manual?

Agilent Technologies M4735A Specifications

General IconGeneral
BrandAgilent Technologies
ModelM4735A
CategoryMedical Equipment
LanguageEnglish

Related product manuals