Optimizing Performance
Creating and Applying User Flatness Correction
Chapter 362
4. Press
Configure Step Array.
This opens a menu for entering the user flatness step array data.
5. Press
Freq Start > 1 > GHz.
6. Press
Freq Stop > 10 > GHz.
7. Press
# of Points > 10 > Enter.
Steps 4, 5, and 6 enter the desired flatness-corrected frequencies into the step array.
8. Press
Return > Load Cal Array From Step Array > Confirm Load From Step Sweep.
This populates the user flatness correction array with the frequency settings defined in the
step array.
9. Press
Amplitude > 0 > dBm.
10.Press
RF On/Off.
This activates the RF output and the RF ON annunciator is displayed on the signal
generator.
Perform the User Flatness Correction
NOTE If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your
power meter does not have a GPIB interface, you can perform the user flatness
correction manually. For instructions, see “Performing the User Flatness
Correction Manually” on page 63.
1. Press
More (1 of 2) > User Flatness > Do Cal.
This creates the user flatness amplitude correction value table entries. The signal
generator enters the user flatness correction routine and a progress bar is shown on the
display.

Table of Contents

Other manuals for Agilent Technologies PSG Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSG Series and is the answer not in the manual?

Agilent Technologies PSG Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSG Series
CategoryPortable Generator
LanguageEnglish

Related product manuals